A. Soleymanpour E. Hamidi Asl, M.A. Nasseri
electroanalysis - , 18, 1598-1604 - May, 2006
Publication year: 2006

ABSTRACT

A new chemically modified carbon paste electrode for cesium(I) ion determination based on potassium zinc hexacyanoferrate (PZHCF) as an ionophore was prepared. The electrode exhibits a Nernstian response for Cs(I) ions over a wide concentration range from 1 × 10–6 to 1 × 10–1 mol·L–1 with a slope of 58 ± 0.5 mV·decade–1. It has a response time of about 35 s and can be used for a period of 3 months with good reproducibility. Detection limit obtained in the optimal conditions was 3 × 10–7 mol·L–1. The potentiometric response is independent of the pH of the solution in the pH range 4.0 – 8.0. The electrode possesses the advantages of low resistance, fast response over a variety of other cations. The proposed electrode is applied as a sensor for the determination of Cs(I) ion concentration in different samples solutions. The results showed a good correlation with the data obtained by atomic absorption spectrometric method