E. Bayat , N. Divani-Vais , M.M. Firoozabadi , N. Ghal-Eh
Rad. Phys. Chem. - 3, 81, 217–220 - January, 2012 - 1390 .
Publication year: 2012

Abstract

The neutron-gamma discrimination capability of the new-type UGLLT liquid scintillator has been compared with NE213 (or its Bicron equivalent, BC501A). An Am-Be combined neutron-gamma isotopic source and a zero-crossing circuitry were used to explore various biases resulting the optimum figure of merits (FOMs) of 1.52 and 1.18 for NE213 and UGLLT at 170 keVee bias, respectively. The study shows that the UGLLT scintillator can be a good alternative for NE213 in pulse-shape discrimination experiments.